Saltar al contenido principal
logo

Dispositivos Soportados

Device TypeModbus TCP (Ethernet)RS485
Inepro PRO380

Cableado

RS485

RS485 Wiring
  • For correct RS485 wiring: Follow the guidelines for RS485 wiring.
  • If the wiring shown in the table below is incorrect, please let us know.
  • There is no general consensus in the industry about the usage of A and B for the RS485 polarity, so it may be counterintuitive and opposite of what you might expect for some devices.
DeviceSofar EMS model OM1Sofar EMS model IG8RS485-USB converterRS485-Ethernet converter
A (pin 22) RS485 ARS485_POSRS485 ATX+
B (pin 23) RS485 BRS485_NEGRS485 BTX-
N/A RS GNDGNDNot availableG
Imagen 1

Configuración

NOTE: RS485 Device Addresses
  • You MUST give each device on the RS485 bus a unique address. Check the manual of the device on how to do this.
  • Use lower addresses first (1, 2, ...) because the Sofar EMS will find them faster!
  • For each device, it is generally recommended to stick with the factory default baud rate, parity, and stop bits. The Sofar EMS will scan on those first.

Consulte el manual del medidor para obtener información más detallada sobre cómo cambiar la dirección Modbus. Por defecto, el medidor está configurado en la dirección 1. Esto se puede cambiar a través de la pantalla del medidor.

  1. Use los botones para ir al Modo de Programación 2.
  2. Presione y mantenga presionado el botón derecho durante 3 segundos para abrir el menú.
  3. Desplácese a la página de LCD ID Modbus/M-bus (addr xxx).
  4. Presione y mantenga presionado el botón derecho durante 5 segundos para ingresar al modo de programación.
  5. El valor comienza a parpadear: seleccione 3 dígitos (Modbus: 001-247). Confirme cada dígito manteniendo presionados ambos botones durante 3 segundos.
  6. El LCD muestra OK cuando se confirma la configuración.
Imagen 1